| Model Number |
626-00-46F |
| Medical Device Problem Codes |
Degraded (1153); Unstable (1667)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Metal Related Pathology (4530)
|
| Date of Event |
07/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
As reported in medwatch: i was in a deep squat and my hip replacement failed and slipped out.I had a bilateral hip replacement with stryker trident psl on 2018.My right hip failed about 3-5 months later cause unknown, right revision on 2019.Left hip failure from a deep squat with the left hip revision surgery on 2022.Additional information reported: patient had a bilateral stryker hip replacement on (b)(6) 2018 with trident product.In (b)(6) 2018, patient noticed instability in right hip.Patient was revised in (b)(6) 2019 due to broken poly liner.Patient mentioned "metal on metal" and alleges high levels of chromium and cobalt.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding abnormal ion level involving an mdm liner was reported.The event was confirmed via clinician review of provided medical records.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: a review of the provided medical information by a clinical consultant indicated: "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck." -product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: it was reported that the patient was revised due to articulation of the ceramic head against the metal mdm liner, damage to the polyethylene adm liner, and elevated metal ion levels.A review of the provided medical information by a clinical consultant indicated: "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck." elevated ion levels that were identified in (b)(6) 2019 is confirmed, but it cannot be confirmed to which device or hip the metal ion levels are attributed.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
As reported in medwatch: i was in a deep squat and my hip replacement failed and slipped out.I had a bilateral hip replacement with stryker trident psl on 2018.My right hip failed about 3-5 months later cause unknown, right revision on 2019.Left hip failure from a deep squat with the left hip revision surgery on 2022.Additional information reported: patient had a bilateral stryker hip replacement on (b)(6), 2018 with trident product.In (b)(6) 2018, patient noticed instability in right hip.Patient was revised in (b)(6) 2019 due to broken poly liner.Patient mentioned "metal on metal" and alleges high levels of chromium and cobalt.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding abnormal ion level involving an mdm liner was reported.The event was confirmed via clinician review of provided medical records.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.Clinician review: a review of the provided medical information by a clinical consultant indicated: "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to articulation of the ceramic head against the metal mdm liner, damage to the polyethylene adm liner, and elevated metal ion levels.A review of the provided medical information by a clinical consultant indicated: "this inquiry concerns patient who underwent bilateral total hip arthroplasties and then subsequently developed disassociation/dislocation of both of his components.There was also elevation of his chrome and cobalt levels.In my opinion a contributing factor is impingement, especially on the right where a notch is seen on the femoral neck.This can account for not only the mechanical failure but also the elevated ion levels.I can confirm that these events occurred because i was able to review the x-rays and operation reports.Regarding the root cause of failure, i cannot determine this for certain.Causes are multifactorial.Dislocation/disassociation can be caused by surgical technique factors, patient activity and lifestyle factors and less likely implant factors.Elevated ion levels in this case are most likely caused by impingement demonstrated clearly on the right with a notch in the femoral neck." elevated ion levels that were identified in july 2019 is confirmed, but it cannot be confirmed to which device or hip the metal ion levels are attributed.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
As reported in medwatch: i was in a deep squat and my hip replacement failed and slipped out.I had a bilateral hip replacement with stryker trident psl on 2018.My right hip failed about 3-5 months later cause unknown, right revision on 2019.Left hip failure from a deep squat with the left hip revision surgery on 2022.Additional information reported: patient had a bilateral stryker hip replacement on (b)(6), 2018 with trident product.In (b)(6) 2018, patient noticed instability in right hip.Patient was revised in (b)(6) 2019 due to broken poly liner.Patient mentioned "metal on metal" and alleges high levels of chromium and cobalt.
|
| |
|
Search Alerts/Recalls
|