Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Medtronic received a report that the device was positioned properly and the distal segment was beginning to flow open.When attempting to move the catheter and ped proximal to the appropriate landing zone, the system jumped proximal to the anz.The physician was unable to navigate the ped back to an appropriate location distal to fully cover the anz neck.The ped was removed and a 4.75x14 was successfully implanted.There was movement during placement.There were not multiple pipeline devices being used when the movement occurred.There was minimal friction during delivery or positioning.The pipeline was not implanted at the intended location.The pipeline missed the landing zone.The device did jump during the deployment.The pipeline was placed at least 3mm past the aneurysm neck on each side.The side branch was not covered by pipeline.The tip of the catheter was moved during the deployment.The device was not used off-label and was prepared as indicated in the ifu.There were no patient symptoms or complications associated with the event.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm in the right ica with a max diameter of 8mm and a 4mm neck diameter.It was noted the patient's vessel tortuosity was severe.Adapt was administered but the pru level was unknown.The angiographic result post procedure showed the ped fully covered the anz with only partial contrast filling present.
|