| Model Number |
20E |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device is not completing the boot up process and has a white display.A white display can be indicative of a device lock up condition.As a result, defibrillation therapy may be unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.Stryker replaced the device's user interface and system pcb assemblies to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.Stryker further evaluated the removed system pcb assembly and wasn't able to determine a further root cause for the reported issue.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device is not completing the boot up process and has a white display.A white display can be indicative of a device lock up condition.As a result, defibrillation therapy may be unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|