• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF MULTIPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number MA60AC
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 11/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
An inventory management specialist reported that during an intraocular lens (iol) implant procedure, the capsular bag was collapsed.Additional information was received stating lens was removed out using cutters and forceps.The surgery was completed on same day with no patient harm.
 
Additional Manufacturer Narrative
Additional information provided in h.3., h.6.And h.10.The reporter indicated the use of a qualified associated cartridge.The use of an unspecified handpiece was indicated.It is unknown if a qualified handpiece was used.A viscoelastic was indicated which is not qualified for use with the lens/cartridge used.The product investigation could not identify a root cause.Information was provided that in the surgeon's opinion, the product did not cause or contribute to the event.No further information was provided.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF MULTIPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14744549
Report Number1119421-2022-01316
Device Sequence Number5465214
Product Code HQL
UDI-Device Identifier00380652552486
UDI-Public00380652552486
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/15/2025
Device Model NumberMA60AC
Device Catalogue NumberMA60AC.180
Device Lot Number15121312
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer08/14/2022
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH II CARTRIDGE B; PROVISC; UNSPECIFIED CUTTERS; UNSPECIFIED FORCEPS; UNSPECIFIED INJECTOR
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-