C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER; TEMPERATURE SENSING PEDIATRIC FOLEY CATHETER Back to Search Results
Model Number 119108
Medical Device Problem Code Partial Blockage (1065)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was no outflow of urine in the foley catheter during the operation and when the passage of distilled water was confirmed with a syringe after removal, it was observed that the lumen of the foley catheter was obstructed.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The device met relevant specifications.The product was not used for patient treatment.The product had not caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), used 3-way temp sensing catheter.Visual inspection of the sample noted the temperature wire was unwinding inside of the catheter.The catheter balloon was inflated with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) successfully with no issue and the catheter rested with no leaks observed.With the syringe attach the balloon deflate passively without issue and the catheter rested with no cuffing or leaks noted.Also noted the drainage valve was also tested, and water flowed through without issue.The photo sample was returned and could not be evaluated for the reported failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review is not required as the reported event is unconfirmed.Corrections: d h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that there was no outflow of urine in the foley catheter during the operation and when the passage of distilled water was confirmed with a syringe after removal, it was observed that the lumen of the foley catheter was obstructed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER
Common Device Name
TEMPERATURE SENSING PEDIATRIC FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14744645
Report Number1018233-2022-04765
Device Sequence Number5176376
Product Code EZL
UDI-Device Identifier00801741045998
UDI-Public(01)00801741045998
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/17/2026
Device Model Number119108
Device Catalogue Number119108
Device Lot NumberNGFV5082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/01/2022
Supplement Date Received by Manufacturer12/06/2022
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-