• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFT015
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A physician reported that during an intraocular lens (iol) implant procedure, when placing the lens and passing it through the cartridge, it fractured and caused inconvenience to the doctor.The lens was removed.Additional information has been requested and received stating the lens was removed by making a larger incision of the cornea.The new lens was implanted the same day.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in h.3., h.6.And h.10.The product was not returned.Information was provided that a qualified company cartridge was used with the lens.A non-qualified viscoelastic and a non-qualified handpiece were indicated.The root cause of the reported complaint was most likely related to a failure to follow the instruction for use (ifu).A non-qualified viscoelastic and handpiece were used.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Company recommends using the qualified company iol delivery system or any other company qualified combination."information was provided that the surgery was "concluded the same day but a lens equal to the fractured one must be obtained".The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14744685
Report Number1119421-2022-01320
Device Sequence Number5176386
Product Code POE
UDI-Device Identifier00380652421201
UDI-Public00380652421201
Combination Product (Y/N)N
Initial Reporter CountryCS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/09/2026
Device Model NumberDFT015
Device Lot Number15201214
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer08/18/2022
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
J AND J VISCOELASTIC AND HANDPIECE; MONARCH III CARTRIDGE D
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
-
-