| Model Number |
21-6500-13 |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pump alarm went off and could not be cleared.The cassette was changed and the alarm was settled.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Premarket (510k) number is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h4: correction: device manufacture date: 21-oct-2020.H6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device in good condition.The customer stated problem was not duplicated.Event history log was reviewed and the error was found multiple times.Replaced possible defective downstream occlusion sensor and the upstream sensor was re-calibrated.The cause of the reported problem could not be determined.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.Therefore, no manufacturing dhr review is needed.
|
| |
|