ARGON MEDICAL DEVICES L-CATH PICC D/L 18GA (3.5F) 1.20MM X 60CM Back to Search Results
Model Number 384467
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
 
Event or Problem Description
¿blue lumen cracked leaking fluids, a new line had to be placed due to patient needs.¿ affect on patient-¿temporary harm, event occurred while in use with patient.¿.
 
Additional Manufacturer Narrative
The sample is indicated as returned.As of the date of this report, the sample has not been evaluated.A follow-up report will be provided once the device has been reviewed.
 
Event or Problem Description
¿blue lumen cracked leaking fluids, a new line had to be placed due to patient needs.¿ affect on patient-¿temporary harm, event occurred while in use with patient.¿.
 
Event or Problem Description
¿blue lumen cracked leaking fluids, a new line had to be placed due to patient needs.¿ affect on patient-¿temporary harm, event occurred while in use with patient.¿.
 
Additional Manufacturer Narrative
A review of the dhr and inspection records was conducted, and no similar concerns were found.One opened catheter was returned for review.Visual inspection confirmed a crack in the luer of the blue lumen that would result in leakage.Several complaints for this part number have previously been received regarding a cracked hub resulting in leakage, and capa 2021-039 was initiated to address this issue.The capa is currently in the implementation phase and will evaluate the corrective action implementation for effectiveness to prevent a recurrence of this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L-CATH PICC D/L 18GA (3.5F) 1.20MM X 60CM
Common Device Name
L-CATH PICC
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key14745081
Report Number0001625425-2022-01033
Device Sequence Number11350632
Product Code LJS
UDI-Device Identifier00886333209866
UDI-Public00886333209866
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K091670
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/06/2025
Device Model Number384467
Device Catalogue Number384467
Device Lot Number11402259
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer06/16/2022
06/16/2022
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date07/26/2022
08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-