| Model Number |
384467 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
|
| |
|
Event or Problem Description
|
|
¿blue lumen cracked leaking fluids, a new line had to be placed due to patient needs.¿
affect on patient-¿temporary harm, event occurred while in use with patient.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample is indicated as returned.As of the date of this report, the sample has not been evaluated.A follow-up report will be provided once the device has been reviewed.
|
| |
|
Event or Problem Description
|
|
¿blue lumen cracked leaking fluids, a new line had to be placed due to patient needs.¿ affect on patient-¿temporary harm, event occurred while in use with patient.¿.
|
| |
|
Event or Problem Description
|
|
¿blue lumen cracked leaking fluids, a new line had to be placed due to patient needs.¿ affect on patient-¿temporary harm, event occurred while in use with patient.¿.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the dhr and inspection records was conducted, and no similar concerns were found.One opened catheter was returned for review.Visual inspection confirmed a crack in the luer of the blue lumen that would result in leakage.Several complaints for this part number have previously been received regarding a cracked hub resulting in leakage, and capa 2021-039 was initiated to address this issue.The capa is currently in the implementation phase and will evaluate the corrective action implementation for effectiveness to prevent a recurrence of this issue.
|
| |
|