• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ID NOW INSTRUMENT; REAL TIME NUCLEIC ACID AMPLIFICATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ID NOW INSTRUMENT; REAL TIME NUCLEIC ACID AMPLIFICATION SYSTEM Back to Search Results
Catalog Number NAT-024
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This investigation is still in progress.Once the investigation is complete a supplemental report will be provided.
 
Event or Problem Description
The customer reported the id now covid-19 instrument to be smoking when the instrument was first turned on.The id now covid-19 instrument was taken out of service.The customer confirmed there was no harm or injury due to the smoking instrument.
 
Additional Manufacturer Narrative
Investigation report: upon receipt at abbott diagnostics scarborough, the instrument was sent to the systems group for further investigation.Systems attempted to power on instrument and observed power was lost immediately after relay energized sending power to base board.Disassembled instrument to further investigate and observed capacitor c122 on the baseboard had failed shorting out 12-volt power supply.C122 is a 16v capacitor in a 12v circuit and is adequately sized to operate under nominal operating conditions however, it is susceptible to damage if there is a voltage transient in the power supply.In a well-designed circuit, capacitors should be rated for a minimum of 2x the intended operating voltage to allow for most transients to pass without damage.The materials surrounding the failing capacitor are either non-flammable (aluminum) or flammability rated with a minimum of 94v-2.These ratings are required for iec 61010-1 compliance.Iec 61010-1 compliance has been commissioned using tuv rhineland and tuv rhineland's safety mark is on the id now ratings label.Cause of customers complaint is a shorted capacitor from a voltage transient above its rated voltage.The release documentation for instrument part number nat024 and instrument serial number (b)(6) was reviewed; the instrument met all release criteria.The reported overheating/smoking issue is anticipated in nature and severity within the product risk file.No new hazard has been identified.In conclusion, the customer's returned instrument was replicated for smoke issue when tested at abbott diagnostics scarborough.A root cause was found to be a failure of the c122 capacitor shorting due to a voltage spike above its rated voltage.The "smoking" reported comes from the capacitor as it burns out.The failed component shorts out the 12-volt dc supply causing the instrument's power pack to shut down as an intended safety feature.This prevents the instrument from turning on.A change order was implemented to increase size of the capacitor to 25v, more than 2x operating voltage to better handle voltage transients.This device was transacted 12dec2018, prior to the capacitor size increase implemented on 09apr2021.The instrument was manufactured prior to the implementation of that change.Ased on the above summary, the investigation is deemed closed.The product will continue to be monitored and tracked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ID NOW INSTRUMENT
Common Device Name
REAL TIME NUCLEIC ACID AMPLIFICATION SYSTEM
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
MDR Report Key14745863
Report Number1221359-2022-03280
Device Sequence Number5465124
Product Code OOI
UDI-Device Identifier00811877010593
UDI-Public0100811877010593
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K173653
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNAT-024
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer07/21/2022
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-