| Model Number |
21-6500-13 |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the use of the product, the "high pressure" alarm went off.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Premarket (510k) number is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: h4: correction: device manufacture date: 08-oct-2020.H6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device in good condition.The customer stated problem was duplicated.Event history log was reviewed and the error was found multiple times.Investigation confirmed that the detection pressure of the downstream occlusion (dso) sensor is out of specification.Replaced defective downstream occlusion sensor and the upstream sensor was re-calibrated.The cause of the reported problem could not be determined.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.Therefore, no manufacturing dhr review is needed.
|
| |
|