• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LASER PROBES; INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER LASER PROBES; INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED Back to Search Results
Catalog Number 8065751113
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that during a combination of cataract and vitrectomy surgery surgeon felt resistance while inserting tip into cannula.The probe was stuck halfway and could not be fully inserted into the cannula.The surgery was completed on the same day and there was no patient harm.
 
Additional Manufacturer Narrative
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.The product was not returned for investigation testing.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The sample was received for testing on this investigation.Initial visual assessment of the returned sample revealed no obvious nonconformity.The returned sample was connected to a calibrated system and was successfully recognized.A visual assessment under a microscope was performed and identified a clear foreign material was present on the sample.A sample of the foreign material was collected, then sent to the lab to be isolated and microscopically analyzed.The analysis of the clear material was found to best match adhesive.An internal investigation into the manufacturing process was conducted.Adhesive is used to bond the nitinol to the cannula and applied at the handpiece/cannula juncture.At no point during the manufacturing process is adhesive applied along the length of the cannula where a trocar may be inserted.Cannula outer diameter is measured on 100% of laser probes to ensure the diameter is within specification prior to release.Thus, based the results of this review of the manufacturing process, how or when the adhesive was introduced onto the cannula is unable to be determined.The root cause of the reported event is attributed to adhesive being present on the cannula.However, how or when the adhesive was introduced onto the cannula remains inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The probe was received for this investigation for testing.A visual assessment of the returned sample revealed no obvious nonconformity.The returned sample was connected to a calibrated system and was successfully recognized.A visual assessment under a microscope was performed and revealed excess adhesive the manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LASER PROBES
Common Device Name
INSTRUMENT, VITREOUS ASPIRATION AND CUTTING, AC-POWERED
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14746849
Report Number2028159-2022-00817
Device Sequence Number5175369
Product Code HQE
UDI-Device Identifier00380657511136
UDI-Public00380657511136
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K946135
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Catalogue Number8065751113
Device Lot Number14431Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer07/26/2022
05/05/2023
06/15/2023
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date08/23/2022
05/30/2023
07/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-