MAUDE Adverse Event Report: SYNTHES PRODUKTIONS GMBH TIBIAL NAIL-ADVANCED / 10MM 375MM / STERILE; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES PRODUKTIONS GMBH TIBIAL NAIL-ADVANCED / 10MM 375MM / STERILE; NAIL, FIXATION, BONE Back to Search Results
Model Number 04.043.245S
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: additional procode: hwc.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during a procedure on (b)(6) 2022, while implanting a 10x375 synthes tibial advance nail, the pin that held the spring in place on the impactor handle had broken.This allowed for the release mechanism to move freely on the impactor.The surgeon noted that the aiming arm was moving independently of the nail, and they attempted to re-tighten the connecting screw, but it was not loose.Upon further inspection of the nail/aiming arm attachment point, it looked like the nail was beginning to splay at its proximal end.The surgeon was able to impact the rest of the nail until it was fully inserted into the patient.The surgery was completed successfully without any other complications.This report is for a tibial nail-advanced / 10mm 375mm / sterile.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h6: a device history record (dhr) review was conducted: part # 04.043.245s.Lot # 679p812.Manufacturing site: jabil bettlach.Release to warehouse date: 22.March.2022.Expiry date: 29.February.2032.A manufacturing record evaluation was performed for the finished device lot and no non-conformance was identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL NAIL-ADVANCED / 10MM 375MM / STERILE
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES PRODUKTIONS GMBH
muracherstrasse 3
bettlach 2544
SZ  2544
Manufacturer (Section G)
WERK BETTLACH
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 
3035526892
MDR Report Key14747733
Report Number9612488-2022-00019
Device Sequence Number20123465
Product Code JDS
UDI-Device Identifier10886982297808
UDI-Public(01)10886982297808
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K201336
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.043.245S
Device Catalogue Number04.043.245S
Device Lot Number679P812
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/23/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DRIVING CAP; UNK - BIOMATERIAL - CEMENT: TRAUMACEM; UNK - INSERTION INSTRUMENTS: CONNECTING: TRAUMA; UNK - NAIL INSERTION HANDLES
Patient SexMale
-
-