Additional Manufacturer Narrative
|
|
The device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure the coil (subject device) prematurely detached.The coil (subject device) was not completely inside the aneurysm.Several stents retrievers and a snare were used as a medical intervention to retrieve the coil (subject device).The patient needed medication to prevent any clot forming around the coil (subject device).The procedure had to be abandon and the patient will come back again for a new attempt.The patient woke up without any deficits.No clinical consequences were reported to the patient due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A2 patient information - age at time of event - added.A2 patient information - age units - added.A3 patient information - gender - added.H4 manufacturing date ¿ added.D4 expiration date - added.Due to the automated mes (manufacturing execution system) system, there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned.
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure the coil (subject device) prematurely detached.The coil (subject device) was not completely inside the aneurysm.Several stents retrievers and a snare were used as a medical intervention to retrieve the coil (subject device).The patient needed medication to prevent any clot forming around the coil (subject device).The procedure had to be abandon and the patient will come back again for a new attempt.The patient woke up without any deficits.No clinical consequences were reported to the patient due to this event.
|
| |
|