MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 STANDARD 7MM X 15CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 STANDARD 7MM X 15CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035467150
Medical Device Problem Codes Expulsion (2933); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during procedure the coil (subject device) prematurely detached.The coil (subject device) was not completely inside the aneurysm.Several stents retrievers and a snare were used as a medical intervention to retrieve the coil (subject device).The patient needed medication to prevent any clot forming around the coil (subject device).The procedure had to be abandon and the patient will come back again for a new attempt.The patient woke up without any deficits.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
A2 patient information - age at time of event - added.A2 patient information - age units - added.A3 patient information - gender - added.H4 manufacturing date ¿ added.D4 expiration date - added.Due to the automated mes (manufacturing execution system) system, there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned.
 
Event or Problem Description
It was reported that during procedure the coil (subject device) prematurely detached.The coil (subject device) was not completely inside the aneurysm.Several stents retrievers and a snare were used as a medical intervention to retrieve the coil (subject device).The patient needed medication to prevent any clot forming around the coil (subject device).The procedure had to be abandon and the patient will come back again for a new attempt.The patient woke up without any deficits.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 STANDARD 7MM X 15CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14748362
Report Number3008881809-2022-00301
Device Sequence Number11944158
Product Code HCG
UDI-Device Identifier07613252192602
UDI-Public07613252192602
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2025
Device Model NumberM0035467150
Device Catalogue NumberM0035467150
Device Lot Number23482218
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer08/10/2022
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexFemale
-
-