STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 44MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Model Number |
623-00-44F |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Ambulation or Postural Difficulties (2544); Metal Related Pathology (4530)
|
| Date of Event |
08/12/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It has been reported "cone wear of the hip stem".Additional information from the surgeon: revision right hip tep via dorsal approach was performed on (b)(6) 2021.Analysis of the joint: metalosis, cone of the hip stem used up, pe inlay damaged, cup trident x3 /58mm fixed, head cocr 44mm/v40cone.Removal of the hip stem via transfemoral approach.The cone of the femoral stem was used up.The femoral head is luxated.Head disassociation from the femoral stem.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding wear involving a trident liner was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: this inquiry reports the revision of a total hip replacement due to head disassociation from the femoral stem with evidence of trunnion deformity and wear.I can confirm that this event took place since i was able since i was able to see the implant removed.Regarding the possible root cause of this event, i cannot determine it with certainty.Causes of trunnion deformity and wear are multifactorial including implant factors, surgical technique factors and patient factors.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the head from the stem.Liner wear/damage was also reported.A review of the provided medical information by a clinical consultant confirmed the disassociation event, but there was insufficient information to confirm wear of the trident liner.Further information such as return of the device/photographs of the device, and the primary/revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|
|
|