Model Number L311
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/13/2022
Event Type  malfunction  
Manufacturer Narrative
Under-insertion of the lead terminal pin into the device header is suspected to be the cause of the malfunction.Note that boston scientific has a vigilant quality monitoring system and we aggressively monitor field performance of all our products.We will continue to monitor for similar events.
 
Event Description
It was reported that during the implant procedure for this pacemaker, the right ventricular (rv) lead showed difficulties to be introduced in the pacemaker header.The rv lead and the pacemaker were successfully implanted but the procedure time was extended due to this issue.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI DR
Type of Device
IMPLANTABLE PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14748681
MDR Text Key294386985
Report Number2124215-2022-22007
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559228
UDI-Public00802526559228
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Model NumberL311
Device Catalogue NumberL311
Device Lot Number604468
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/16/2022
Initial Date FDA Received06/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age84 YR
Patient SexFemale
-
-