• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD MAXON; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD MAXON; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number 8886605941
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, on an open abdominal gastric resection, during closure, prior to using the device, the thread was found to have come off the needle.Another device was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted one suture and one needle detached.The needle was returned disengaged from the suture.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.It was reported that the needle unexpectedly separated from the thread.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXON
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key14748865
Report Number9612501-2022-00997
Device Sequence Number17533411
Product Code GAM
UDI-Device Identifier10884523001921
UDI-Public10884523001921
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K990951
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model Number8886605941
Device Catalogue Number8886605941
Device Lot NumberD0G0013FY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2022
Supplement Date Received by Manufacturer07/29/2022
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-