• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. EVOLUTION® POROUS KEEL PRIMARY 15MM X SIZE 5/6; KNEE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. EVOLUTION® POROUS KEEL PRIMARY 15MM X SIZE 5/6; KNEE COMPONENT Back to Search Results
Model Number ETPK1556
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event or Problem Description
Allegedly, 7 months post op had increased weight bearing pain.44 months post op- no complaint just visit to doctor but can see some loosening on (b)(6) 2022 patient was in pain, hyperflex occurred in show due to other issues components have not been revised.This was a prophecy case (b)(4).
 
Additional Manufacturer Narrative
Additional information received on 6/29/2022.This manufacturer report number (b)(4) is considered no longer reportable since it has been indicated that the adverse event was caused by a traumatic event (accident with a scooter that led to a bent/broken tibia base).The incident is related to trauma and not related to a device failure.Please void this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION® POROUS KEEL PRIMARY 15MM X SIZE 5/6
Common Device Name
KNEE COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key14750015
Report Number3010536692-2022-00227
Device Sequence Number2263952
Product Code MBH
UDI-Device IdentifierM684ETPK15561
UDI-PublicM684ETPK15561
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberETPK1556
Device Catalogue NumberETPK1556
Device Lot Number1753594
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/03/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date06/20/2022
Supplement Report FDA Received Date08/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
-
-