| Model Number |
MI1250 SYNCHRONY 2 PIN |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
The clinic reportedly had difficulties with the intraoperative measurements.After troubleshooting, the user was re-implanted with a backup device.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the device investigation revealed a technical implant failure i.E.Inadequate inclination of the circuitry substrate, which was likely caused during manufacturing process.The recipient was re-implanted with a back-up device within the same surgical session.This is a final report.
|
| |
|
Event or Problem Description
|
|
The clinic reportedly had difficulties with the intraoperative measurements.The device was indicative of a device malfunction.After troubleshooting, the affected device was removed and the recipient was finally implanted with a backup device.The wound had already been closed before removing the concerned device; the surgery was prolonged.However, all the procedure was performed under one session of anaesthesia.
|
| |
|