| Model Number |
RS271A30 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Superficial (First Degree) Burn (2685); Partial thickness (Second Degree) Burn (2694)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On may 25th, 2022 we have been informed about an incident involving a dispersive electrode.A cardiovascular procedure was performed at (b)(6) hospital in the us.A megadyne dispersive electrode catalog number 0855c (our model rs271a30) and ft10 and force fx generators generator were used.The initial user report stated: "historical perspective from surgeon mid (b)(6) and early (b)(6).4 separate patients with wounds after the or.Suspicious for burns.Areas include where the pad was or where the cords were.No similarities in procedures, or staff, generators, surgeons, different lot numbers, etc.No pads were saved for analysis.The surgeon could not speak if a burn specialist has confirmed if this was thermal or chemical related yet." it was also stated in the form: "a single photo associated with (b)(4) is evaluated.The photo, labelled as right back, was taken on 4/14/22.A slightly irregular oval area of skin lesion is visualized.The area skin shows mix of region of light darkness and patch of redness.The lesion appears limited in the superficial layer of skin with no swollen or blister visible.The skin lesion is a first degree burn appearance but no underlying causes can be determined from the limited information available." "the wire came under the arm and was plugged in the esu and that is where the patient got burned.The patient with burn in the armpit.The file with the back burns the surgeon wondered if there was a reaction to the sticky pad.All photos were taken 3-4 days later.All burns were noticed in post op.The rep went into the or to see if the nurses were putting the sticky pads on correctly and they were.1 esu at the foot and 1 esu at the head.Pad was placed on the flank and the wire came under the arm to be plugged into the esu.Most of time uses 2 pads for 2 different generators.Ft10 and force fx generators used.No alarms were heard during procedure.Different generators were used.All different or rooms.All burns were under the cord or pad site.Came from multiple lot numbers.Rep does not know about skin prep for the patient." on june 15th, 2022 we have received a picture showing an approximately 13cm long and 4cm wide burn.In the photo a part of the burn site appears pink (3x4 cm) with dark areas extending beyond the pink, probably blistered.No further information was provided despite repeated requests.
|
| |
|
Additional Manufacturer Narrative
|
|
As neither a lot number nor samples have been made available to us, no analyses of dispersive electrodes could be performed.We have requested additional information and the involved device in the incident but have not yet received either.We have requested further information but so far only a photo of the burn has been made available to us.The photo provided will be reviewed by a dermatologist.We will provide a follow-up report once we receive further information.
|
| |
|
Event or Problem Description
|
|
On may 25th, 2022 we have been informed about an incident involving a dispersive electrode.A cardiovascular procedure was performed at adventhealth orlando hospital in the us.A megadyne dispersive electrode catalog number 0855c (our model rs271a30) and ft10 and force fx generators generator were used.The initial user report stated: "historical perspective from surgeon mid april and early may.4 separate patients with wounds after the or.Suspicious for burns.Areas include where the pad was or where the cords were.No similarities in procedures, or staff, generators, surgeons, different lot numbers, etc.No pads were saved for analysis.The surgeon could not speak if a burn specialist has confirmed if this was thermal or chemical related yet." it was also stated in the form: "a single photo associated with (b)(4) is evaluated.The photo, labelled as right back, was taken on (b)(6) 2022.A slightly irregular oval area of skin lesion is visualized.The area skin shows mix of region of light darkness and patch of redness.The lesion appears limited in the superficial layer of skin with no swollen or blister visible.The skin lesion is a first degree burn appearance but no underlying causes can be determined from the limited information available." "the wire came under the arm and was plugged in the esu and that is where the patient got burned.The patient with burn in the armpit.The file with the back burns the surgeon wondered if there was a reaction to the sticky pad.All photos were taken 3-4 days later.All burns were noticed in post op.The rep went into the or to see if the nurses were putting the sticky pads on correctly and they were.1 esu at the foot and 1 esu at the head.Pad was placed on the flank and the wire came under the arm to be plugged into the esu.Most of time uses 2 pads for 2 different generators.Ft10 and force fx generators used.No alarms were heard during procedure.Different generators were used.All different or rooms.All burns were under the cord or pad site.Came from multiple lot numbers.Rep does not know about skin prep for the patient." on june 15th, 2022 we have received a picture showing an approximately 13cm long and 4cm wide burn.In the photo a part of the burn site appears pink (3x4 cm) with dark areas extending beyond the pink, probably blistered.No further information was provided despite repeated requests.
|
| |
|
Additional Manufacturer Narrative
|
|
As neither a lot number nor samples have been made available to us, no analyses of dispersive electrodes could be performed.We have requested additional information and the involved device in the incident but have not yet received either.We have repeatedly requested for customer samples, the completed questionnaire, and further information.On july 12th, we have been informed that "no additional information received" and in a further communication it was specified that "this file has been completed and is in closed status".No conclusion can be drawn what might have caused the claimed product problem.As no further information was available, we therefore consider the investigation and the report closed.
|
| |
|