WRIGHTS LANE SYNTHES USA PRODUCTS LLC HAERLOCK 2 IMPL KIT 12X5/0 DEG/SMALL; PIN, FIXATION, SMOOTH
|
Back to Search Results |
|
| Model Number |
HL2S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, hammer lock 2 implants were broken without the package being opened.There was no patient involvement.This report is for one (1) haerlock 2 impl kit 12x5/0 deg/small this is report 1 of 2 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: e1 h6: device photo investigation: the photo was returned to depuy synthes for evaluation.Visual analysis of the photo revealed that the haerlock 2 impl kit 12x5/0 deg/small was broken from the inserter.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the haerlock 2 impl kit 12x5/0 deg/small.Root cause for this failure mode has been determined to be shipping/handling processes and design issues.Capa was opened to drive this investigation.Corrective actions (design and packaging design changes) have been implemented, released january 11, 2018.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|