WRIGHTS LANE SYNTHES USA PRODUCTS LLC HAERLOCK 2 IMPL KIT 15X7/0 DEG/LRG; PIN, FIXATION, SMOOTH
|
Back to Search Results |
|
| Model Number |
HL2L |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, hammer lock 2 implants are broken without the package being opened.There was no patient involvement.This report is for one (1) haerlock 2 impl kit 15x7/0 deg/lrg this is report 2 of 2 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: e1 h6: photo investigation: the photo was returned to depuy synthes for evaluation.Visual analysis of the photo revealed that the haerlock 2 impl kit 15x7/0 deg/lrg was broken from the inserter.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the haerlock 2 impl kit 15x7/0 deg/lrg.This is a known issue with hl2 implants.Process car 1110 was opened in bme's old qms system (qcbd) to investigate this issue with hl2l and hl2la.After investigation, it was determined that the root cause was a molding issue from the provider (protolabs).A new mold from the provider was obtained as corrective action.The lot for the devices involved in this complaint.It was later determined that shipping was a contributing factor for this type of occurrences under capa.Corrective actions (design and packaging design changes) have been implemented, released (b)(6) 2018.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|