| Model Number |
00055445010 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The manufacturer received a complaint indicating a gem 5000 sensor and reagent cartridge was received with mold on it.The customer reports upon opening the packaging, mold was observed on the exterior of the cartridge case and/or labeling.
|
| |
|
Additional Manufacturer Narrative
|
|
Instrumentation laboratory conducted an investigation that included a review of the customer-provided photographs demonstrating small amounts of mold on the gem 5000 sensor and reagent cartridge.This event may have occurred due to a possibly broken or incomplete seal of an internal reagent bag.Exposure to environmental molds can cause local skin, eye, or mucosal membrane irritation.Standard laboratory practice is to handle all reagents/cartridges wearing appropriate personal protective equipment, thus limiting exposure to mold or other contaminants.Thus, it is improbable that harm would occur due to the event.No remedial action is required.
|
| |
|