| Model Number |
CSE-EW-UK |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A haemonetics field service engineer evaluated the data log files from the cell saver elite system used in the procedure.Data showed that the device functioned correctly as there were no early trip to wash events caused by the device.A dhr review was performed and the product met all specifications during final manufacturing tests.There are no corrective actions or non-conformances related to this event.The disposables used during the procedure were not returned for evaluation.There is no evidence to suggest that the event was related to the device used during the procedure.There was no report of patient harm reported in this event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, haemonetics was notified of "failure to empty 3rd bowl into rbc bag x 2, new bowl set loaded.4th bowl with new bowl set also failed to empty.Blood from both bowls discarded.Donor transfusion received / quantity: none received (confirmed (b)(6) 2021).Cell salvage blood re-infused / volume: 221 ml re-infused (x 2 full bowls discarded) - approx another 225 ml" during a procedure using a cell saver® elite® plus autotransfusion system and cell saver® elite set-125m in the united kingdom.It was decided to communicate an initial mdr report due to potential procedure delay.No harm to the patient was reported in this event.
|
| |
|