| Model Number |
VAMF4040C200TU |
| Medical Device Problem Codes |
Malposition of Device (2616); Material Deformation (2976); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A valiant captivia stent graft was implanted while 3 other valiant captivia stent grafts were intended to be implanted in the endovascular treatment of a 80mm descending thoracic aortic aneurysm via percutaneous approach.It was reported during the procedure, the physician experienced problems getting the device to track to the intended landing zone. the patient has a bovine arch and the intended landing zone was at the origin of the brachiocephalic artery.Multiple attempts were made to pass the device without any luck.The device would make it to the left subclavian but would never pass around the arch. after multiple attempts the physician then came from the right arm to body floss and still had the same issue not able to get the device to trackaround the arch.At this point the physician deployed a valiant captivia well below the intended landing zone to see if that would provide more support to get a new device around the arch to the intended landing zone and the vamf4040c200tu lot #v30682555 was then successfully implanted. 3 other valiant captivia's (#v30682553 #v30731110t#v30580527) were opened and all 3 kinked trying to get them around the arch.Each device had the same problem kinking going through the distal thoracic aorta.Each device kinked in roughly the same spot. after kinking three devices the physician called the non mdt sales rep who brought a competitor stent graft in to try to complete the procedure. they placed the non mdt device in and it tracked with no problems to get to the intended landing zone.The graft was deployed and ballooned with no issues.The procedure was then completed and the physician then closed all the access sites.As per the physician the cause of the event was anatomy. no additional clinical sequalae were provided and the patient is fine.The patient is awake and moving all extremities.No signs of stroke due to all manipulations in the arch.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|