MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS ADD-ON BURETTE SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS ADD-ON BURETTE SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 82113
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd alaris add-on burette set experienced leakage.The following information was provided by the initial reporter: add-on burette set for iv measurement appear to be leaking from the base of the set.
 
Additional Manufacturer Narrative
H6: investigation summary: the customer reported there was leaking from the bottom of the buretrol, and returned one unused 82113.The burette set was tested for leakage and the complaint was verified.The buretrol leaked at the connection of the clear plastic and the white bottom.It was also easy to squeeze the buretrol and pull the clear plastic out of the white.Analysis under the microscope found glue to be applied to only one half of the buretrol and bottom.A device history record review for model 82113 lot number 21119767 was performed.The search showed that a total of (b)(4) units lot number was built on 23nov2021.There was a quality notification issued for the failure mode when the leak test was not passed, and manufacturing began working on the issue.Manufacturing found the root cause to be partial gumming on the ends of the solvent dispenser.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd alaris add-on burette set experienced leakage.The following information was provided by the initial reporter: add-on burette set for iv measurement appear to be leaking from the base of the set.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS ADD-ON BURETTE SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14754176
Report Number9616066-2022-00741
Device Sequence Number11086234
Product Code FPA
UDI-Device Identifier37613203009622
UDI-Public37613203009622
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K780130
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/18/2024
Device Model Number82113
Device Catalogue Number82113
Device Lot Number21119767
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer08/02/2022
Initial Report FDA Received Date06/21/2022
Supplement Report FDA Received Date08/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-