| Catalog Number |
MC1005T |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a roi-c plate would not fit in the cage intra-operatively.There were no reported patient impacts.Device evaluation by the manufacturer found the plate was bent.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection revealed that the device has a bent arm and minor dings/gouges.Root cause: root cause was unable to be determined.It is possible to occur from hard (sclerotic) bone and not using the starter awl, skipping impactor #1 and using only impactor #2, or interference with an adjacent construct (typically a caspar pin).Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|