• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP AUG MINI BSPLT W TPR MD; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP AUG MINI BSPLT W TPR MD; PROSTHESIS, SHOULDER Back to Search Results
Catalog Number 110032420
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Bacterial Infection (1735); Pain (1994); Osteopenia/ Osteoporosis (2651)
Date of Event 01/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Common device name: phx.Concomitant medical products: comp rvs cntrl 6.5x30mm st/rst cat#115396 lot# 576760; comp lk scr 3.5hex 4.75x30 st cat# 180553 lot# 246200; comp lk scr 3.5hex 4.75x35 st cat# 180554 lot# 698810; comp lk scr 3.5hex 4.75x25 st cat# 180552 lot# 591530; comp lk scr 3.5hex 4.75x15 st cat# 180550 lot# 419550; unk humeral stem cat# unk lot# unk; cr vivacit-e 40mm brng +3 cat#110031428 lot#64525936; cr 40mm glenosphere +3mm cocr cat# 110030777 lot# 64328699; mini tray std cocr +0 offset cat# 110031399 lot# 64957156.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 01458; 0001825034 - 2022 - 01459; 0001825034 - 2022 - 01460; 0001825034 - 2022 - 01461; 0001825034 - 2022 - 01462; 0001825034 - 2022 - 01463.
 
Event or Problem Description
It was reported that a patient underwent a revision procedure approximately 10 months postimplantation due to infection, persistent pain and loosening of hardware.All components were removed and an antibiotic spacer was placed.Significant discolored bio-film type tissue was present upon removal of the humeral tray.Humeral stem was removed without significant bone loss, glenoid removed without complication.The baseplate was grossly loose prior to removal, with significant bone loss, soft tissue debridement and bio-film tissue present.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, therefore implanted products are not identified as the source or contributing to the reported infection.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP AUG MINI BSPLT W TPR MD
Common Device Name
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14754313
Report Number0001825034-2022-01457
Device Sequence Number14991011
Product Code PHX
UDI-Device Identifier00880304993242
UDI-Public(01)00880304993242(17)230625(10)64089391
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K172502
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2023
Device Catalogue Number110032420
Device Lot Number64089391
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2022
Supplement Date Received by Manufacturer07/07/2022
Initial Report FDA Received Date06/21/2022
Supplement Report FDA Received Date07/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/25/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight103 KG
-
-