| Model Number |
AHX500S11 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient alleged seeing visualization of particles in the device.There was no report of patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously reported an allegation of an issue related to a cpap device's sound abatement foam.The manufacturer received information alleged visualization of particles.There was no report of serious or permanent harm or injury. the device was returned to the manufacturer's service center for further evaluation.The device was evaluated.There was no mention of visual findings to the external part of the device.The internal aspect of the device was inspected.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the manufacturer.There was 0 errors found. the manufacturer concludes that they could not confirm the customer's allegation and there was no visible foam degradation and the device was scrapped due to age.Section d8, d9 and h6 updated in this report.
|
| |
|