Model Number MX9059
Medical Device Problem Code Contamination of Device Ingredient or Reagent (2901)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there were multiple occurrences of the bd extension set maxplus¿ experiencing foreign matter.The following information was provided by the initial reporter: it was found that there is yellowing component.
 
Additional Manufacturer Narrative
Investigation summary: twenty samples were submitted for quality investigation.The customer complaint of foreign matter was verified by visual inspection.Evaluation of the samples submitted identified a yellow discoloration in the tubing and extension set connectors.No other issues were found during the investigation.A device history record review for model mx9059, lot number 21065711 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 28jun2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A device history record review for model mx9059, lot number 21045469 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 06apr2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A review of the sterilization record for item #mx9059, lot 21065711 shows that the extension sets were sterilized in (b)(6) 2021.A review of the sterilization record for item #mx9059, lot 21045469 shows that the extension sets were sterilized in (b)(6) 2021.The root cause for the issue seen in this complaint is a result in the sterilization process.Previous complaints have seen a similar issue with discoloration of extension set tubing, with the tinting of the tubing increasing over time.Previous investigations have shown that the discoloration does not affect the performance or sterility of the product.Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 21045469, medical device expiration date: 06-apr-2026, device manufacture date: 05-apr-2021; medical device lot #: 21065711, medical device expiration date: 28-jun-2026, device manufacture date: 09-jun-2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD EXTENSION SET MAXPLUS¿
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14754884
Report Number9616066-2022-00743
Device Sequence Number1334348
Product Code FPA
UDI-Device Identifier50885403237582
UDI-Public50885403237582
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K051499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberMX9059
Device Catalogue NumberMX9059
Device Lot NumberSEE H10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Initial Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-