| Model Number |
0700410105 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Alleged failure: surgeons complains of grainy video and not optimal focus.Surgeon cannot get optimal focus using the coupler.[update - no loss of image, replacements used to complete procedure].The failure(s) identified in the investigation is consistent with the complaint record.Root cause: short in cable due to the sterilization cycles and/or bending, kinking, high force impact or other damage to the cable.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.The device manufacturer date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|