INTEGRA LIFESCIENCES SWITZERLAND SAR NEUROMONITOR BASIC KIT; ICP MICROSENSORS Back to Search Results
Model Number 82-6631
Medical Device Problem Code Device Damaged Prior to Receipt by User (1570)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
A facility reported a box of a neuromonitor basic kit (id 826631) was crushed before the procedure.The procedure was completed with a replacement product available.No surgical delay and adverse consequences to the patient were observed.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The neuromonitor basic kit (id 826631) was returned for evaluation.Device history record (dhr) - the product 826631 for lot 4025627 and the lot met specifications when released.Failure analysis - it cannot be confirmed the damages occurred at the integra distribution center.It cannot be excluded the damages occurred during transit from distribution center to customer.Root cause cannot be determined.Based on information provided it cannot be determined if this was a packing error at the distribution center or if it occurred during transit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROMONITOR BASIC KIT
Common Device Name
ICP MICROSENSORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14755503
Report Number3013886523-2022-00282
Device Sequence Number6153522
Product Code GWM
UDI-Device Identifier20886704042317
UDI-Public20886704042317
Combination Product (Y/N)N
PMA/510(K) Number
K914479
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model Number82-6631
Device Catalogue Number826631
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/03/2022
Supplement Date Received by Manufacturer03/30/2023
Initial Report FDA Received Date06/21/2022
Supplement Report FDA Received Date04/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-