| Model Number |
750042 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Internal Organ Perforation (1987)
|
| Date of Event |
06/10/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The perforation was detected via cystoscopy.Light bleeding was observed.Contrast medica appeared on the uromedica u-channel sheath following the perforation.The physician indicated that he dilated with the blunt trocar, and felt resistance build up and then give way.Uromedica complaint#: (b)(4).
|
| |
|
Event or Problem Description
|
|
Interoperative bladder perforation (patient's left side) during a proact implantation procedure.The bilaterial implantation procedure was completed successfully.Perforation treated with temporary catheterization.
|
| |
|
Search Alerts/Recalls
|