SYNTHES PRODUKTIONS GMBH TIBIAL NAIL-ADVANCED / 9MM 360MM / STERILE; NAIL, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
04.043.140S |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional procode: hwc.Complainant part is not expected to be returned for manufacturer review/ investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, during a routine tibial nailing of a midshaft tibia fracture, the surgeon experienced a problem where the aiming arm for the proximal interlocks did not line up with the most distal hole in the nail, and the nail obstructed the path of the drill bit.The patient¿s knee was bumped up to remove tension at the nail/insertion handle junction, and the connection point was inspected and confirmed to be tight, but the issue was not resolved.The aiming arm was removed and the procedure was completed using a free-hand perfect circle technique.No patient impact.This report is for a tibial nail-advanced / 9mm 360mm / sterile.This is report 1 of 3 for (b)(4).
|
| |
|
Event or Problem Description
|
|
It was also reported that there was a ten (10) minutes of surgical delay.No fragments left in patient.Surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|