BECTON DICKINSON MEDICAL SYSTEMS BD INTEGRA SYRINGE; SAFETY SYRINGE W/ NEEDLE Back to Search Results
Catalog Number 305283
Medical Device Problem Codes Crack (1135); Leak/Splash (1354); Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd integra syringe experienced device damage while still considered operable, and leakage.The following information was provided by the initial reporter: upon minimal pressing of the pestle, the syringe broke and the medication flowed out at the top and bottom.Subsequently, the system was also defective.- what medication/fluid was used: temesta 4mg - did a serious injury occur? no - did any medical action have to take place as a result of this incident?: no.
 
Additional Manufacturer Narrative
Device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd integra syringe experienced device damage while still considered operable, and leakage.The following information was provided by the initial reporter: upon minimal pressing of the pestle, the syringe broke and the medication flowed out at the top and bottom.Subsequently, the system was also defective.- what medication/fluid was used: temesta 4mg, - did a serious injury occur? no, - did any medical action have to take place as a result of this incident?: no.
 
Additional Manufacturer Narrative
H.6 imdrf annex a code(s): a0510 -retraction problem, a0504 -leak/splash.H.6 investigation summary: one loose sample of a 3ml integra syringe (p/n 305283) was received.A visual evaluation performed, no barrel cracked or damaged observed, however; the cutter had cut thru the stopper causing the fluid to leak out of the stopper, but the needle was still inside the needle hub as it had not been activated.The condition observed is non-conforming per product specification.Potential root cause for the premature retraction and needle retraction failure defects are associated with the assembly process.The batch number is unknown, therefore defective rate cannot be identified and corrective actions are not necessary.The batch number is unknown, therefore device history record review (dhr) or quality notification review (qn) could not be performed.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Potential root cause for the premature retraction and retraction failure defects are associated with the assembly process.
 
Additional Manufacturer Narrative
The following information has been corrected: h.6 imdrf annex a code(s): a0510 -retraction problem, a0504 -leak/splash.
 
Event or Problem Description
It was reported that the bd integra syringe experienced device damage while still considered operable, and leakage.The following information was provided by the initial reporter: upon minimal pressing of the pestle, the syringe broke and the medication flowed out at the top and bottom.Subsequently, the system was also defective.- what medication/fluid was used: temesta 4mg, - did a serious injury occur? no, - did any medical action have to take place as a result of this incident?: no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTEGRA SYRINGE
Common Device Name
SAFETY SYRINGE W/ NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14756783
Report Number1213809-2022-00348
Device Sequence Number1348375
Product Code MEG
UDI-Device Identifier00382903052837
UDI-Public00382903052837
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K011103
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number305283
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer07/06/2022
08/03/2022
Initial Report FDA Received Date06/21/2022
Supplement Report FDA Received Date08/03/2022
08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-