• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION Back to Search Results
Model Number 560-011
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
A lead extraction procedure commenced to remove a right atrial (ra) and a right ventricular (rv) lead due to bacteremia.The physician chose a spectranetics tightrail sub-c rotating dilator sheath to aid in lead removal.During the procedure, the tightrail device stopped working.There was a difference in the sound it made as the physician squeezed the handle, as if it had broken inside and was no longer turning, which was confirmed under fluoroscopy.After use, it was observed the device''s shaft was very loose and flimsy.A new device was opened and the procedure was completed with no reported patient harm.The tightrail device was received in two pieces on (b)(6) 2022.The handle and the shaft were separate, which would prevent the blades (located at the distal tip of the device''s shaft) from actuating, since actuation of the blades originates in the device''s handle.During device evaluation on (b)(6) 2022, it was discovered that the blades of the tightrail device were not retracted and were in an extended position, past the outer band of the device.This report captures the tightrail device returned with the separated shaft and extended blades, potential for harm.
 
Additional Manufacturer Narrative
The device was evaluated by a cross functional team.The handle and shaft were returned in two pieces, with the inner and outer coils broken under the strain relief.Severe biologics were seen throughout the device.Observing the distal tip of the shaft, the blades were found to be in an extended position.With the handle halves separated, the device showed heavy to normal use.Due to the significant amount of biologics seen on the inside of the device, the barrel was unable to rotate within the sleeve.After the device was soaked and ultrasonically cleaned, the barrel revealed severe marring on the barrel on the home track return path and slight marring on the far track return path.This marring caused the barrel and sleeve to fail the drop test.Manual actuation could not be tested due to the shaft being broken into two pieces.Additional analysis showed evidence of a uniform directional ductal tensile break, indicative of stress being placed on the device.There was no evidence of a design, production or use related failure.It was concluded that the broken shaft was caused by exceeding the design limits of the tensile strength of the device during use.However, it could not be determined how or when the limits of the tensile strength were exceeded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH
Common Device Name
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key14758298
Report Number1721279-2022-00117
Device Sequence Number1348414
Product Code DRE
UDI-Device Identifier00813132022792
UDI-Public(01)00813132022792(17)230901(10)FRP21J01A
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K161333
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2023
Device Model Number560-011
Device Catalogue Number560-011
Device Lot NumberFRP21J01A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Initial Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MEDTRONIC 5086 RA PACING LEAD; MEDTRONIC 5086 RV PACING LEAD; TRACTION PLATFORM, MANUFACTURER UNK
Patient Age52 YR
Patient SexMale
-
-