• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH TIBIAL NAIL-ADVANCED / 8MM 285MM / STERILE; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TIBIAL NAIL-ADVANCED / 8MM 285MM / STERILE; NAIL, FIXATION, BONE Back to Search Results
Model Number 04.043.015S
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from japan reports an event as follows: it was reported that on (b)(6) 2022, during a procedure for orif with a tna for the fracture of the diaphyseal tibia, the surgeon inserted the nail to the optimal depth to begin drilling for a proximal locking screw.The drill bit interfered with the nail hole while drilling, in both ml holes.After investigating, it was discovered that the interference occurred behind the hole, so the surgeon drilled again while adjusting by hand.The drill bit interfered with the nail hole, but the surgeon was able to drill and insert screws.While removing the insertion handle, the connecting screw came off with a rasping sound and the surgeon concluded that nail deflection had occurred.The surgery was completed successfully with less than a 30 minutes delay.This report is for a tibial nail-advanced / 8mm 285mm / sterile.This is report 1 of 2 for (b)(6).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional procode: hwc.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL NAIL-ADVANCED / 8MM 285MM / STERILE
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 
3035526892
MDR Report Key14758462
Report Number8030965-2022-04205
Device Sequence Number12288782
Product Code JDS
UDI-Device Identifier10886982297440
UDI-Public(01)10886982297440
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K201336
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.043.015S
Device Catalogue Number04.043.015S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/25/2022
Initial Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AIMING ARM/ RADIOLUCENT; DRIVING CAP; INSERTION HANDLE/ RADIOLUCENT MEDIUM; UNK - DRILL BITS: TRAUMA; UNK - IMPACTION INSTRUMENTS: HAMMER/MALLET: TRAUMA; UNK - INSERTION INST: COUPLING SCREW: TR; UNK - SCREWS: NAIL PROXIMAL LOCKING
Patient Age55 YR
Patient SexUnknown
-
-