Model Number 40000-07T
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd alaris texium secondary set experienced leakage which lead to patient not receiving the entire dose of medication.The following information was provided by the initial reporter: chemotherapy dripping from between the male leur and the texium end.This resulted in a chemo spill and patient did not receive the entire dose of the medication - oxaliplatin.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 22-jun-2022.H6: investigation summary.The customer reported dripping between the texium and the male luer.The sample came with these two attached, and when the set was primed, there was no leakage.However, the connection was not normal.The texium was extremely tight in the luer, and they could not be separated.The texium was also not inserted fully into the male luer threads, it appeared to be threaded incorrectly.Perhaps the medication or other solution crusted around the connection and prevented further leakage.The complaint will be verified as connection issues.The sample was sent to the manufacturing site for further investigation and they found the root cause to be the torque fixture for the assembly of the connection.A quality alert was issued to the assembly personnel to address the issue.A device history record review for model 40000-07t lot number 22019006 was performed.
 
Event or Problem Description
It was reported that the bd alaris texium secondary set experienced leakage which lead to patient not receiving the entire dose of medication.The following information was provided by the initial reporter: chemotherapy dripping from between the male leur and the texium end.This resulted in a chemo spill and patient did not receive the entire dose of the medication - oxaliplatin.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS TEXIUM SECONDARY SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14758548
Report Number9616066-2022-00748
Device Sequence Number6022863
Product Code FPA
UDI-Device Identifier50885403225626
UDI-Public50885403225626
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K790582
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/03/2025
Device Model Number40000-07T
Device Catalogue Number40000-07T
Device Lot Number22019006
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer11/09/2022
Initial Report FDA Received Date06/21/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-