SYNTHES GMBH 3.5MM TI LCP® METAPHYSEAL PLATE 9 HOLES; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, METAL COMPOSITE
|
Back to Search Results |
|
| Model Number |
423.409 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: event occurred at an unknown date in 2022, after (b)(6) of 2022.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from japan reports an event as follows: it was reported that on (b)(6) 2021, a patient underwent orif surgery for a tibial shaft fracture with the plate in question.The surgery was completed successfully.It was confirmed via x-ray that no malfunction had occurred until (b)(6) 2022.On an unknown date in (b)(6) 2022, the breakage of the plate was confirmed and removal surgery was performed on (b)(6) 2022.The removal surgery was completed successfully without any surgical delay.This report is for a 3.5mm ti lcp® metaphyseal plate 9 holes.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9, g3: device returned 20-july-2022 g6: follow up #1 h2, h3, h6: device evaluated.Part # 423.409s lot # 1l78617 manufacturing site: (b)(4) release to warehouse date: 17 oct2018 a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that (lcp metaphyspl 3.5 9ho l125 ti) has broken in two pieces.The broken fragments were returned.No other problem was reported.A dimensional inspection was not performed for the lcp metaphyspl 3.5 9ho l125 ti due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the lcp metaphyspl 3.5 9ho l125 ti would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|