AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEM FOR NEONATES / PEDIATRICS, 5 F, Y-ADAPTER; VAP CLOSED SUCTION CATHETERS & ACCESSORIES
|
Back to Search Results |
|
| Model Number |
195-5 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Codes |
Bradycardia (1751); Low Oxygen Saturation (2477)
|
| Date of Event |
05/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 20-jun-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported 'during tracheal aspiration of the child with a closed system, (aspiration probe in the process of descent into the intubation tube), the plastic part surrounding the tube began to swell and the child to bradycardia and desaturate.' additional information received 01-jun-2022 stated the patient was 5-days post-op for ductus arteriosus surgery, ventilated in cpap with caffeine dopram support.The patient was parenterally and enterally fed.The patient still has numerous episodes of bradycardia.Additional information received 10-jun-2022 stated at the time of the event the caregiver performing aspiration managed the destabilization of the patient.The suction catheter was replaced with a new one that did not cause any problems in use.Additionally, it was noted surgery, ventilation, and diet were part of the patient's health condition but the prolonged bradycardia was due to the use of the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction d4: 30147513.The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 17-aug-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
All information reasonably known as of 27-jul-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for the reported lot number, 30147513, was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.The returned product was evaluated and no issues were observed on device.The root cause could not be determined since the reported failure was not reproduced in lab during sample evaluation.All information reasonably known as of 28-sep-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|