MAUDE Adverse Event Report: AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEM FOR NEONATES / PEDIATRICS, 5 F, Y-ADAPTER; VAP CLOSED SUCTION CATHETERS & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEM FOR NEONATES / PEDIATRICS, 5 F, Y-ADAPTER; VAP CLOSED SUCTION CATHETERS & ACCESSORIES Back to Search Results
Model Number 195-5
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Codes Bradycardia (1751); Low Oxygen Saturation (2477)
Date of Event 05/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 20-jun-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Event or Problem Description
It was reported 'during tracheal aspiration of the child with a closed system, (aspiration probe in the process of descent into the intubation tube), the plastic part surrounding the tube began to swell and the child to bradycardia and desaturate.' additional information received 01-jun-2022 stated the patient was 5-days post-op for ductus arteriosus surgery, ventilated in cpap with caffeine dopram support.The patient was parenterally and enterally fed.The patient still has numerous episodes of bradycardia.Additional information received 10-jun-2022 stated at the time of the event the caregiver performing aspiration managed the destabilization of the patient.The suction catheter was replaced with a new one that did not cause any problems in use.Additionally, it was noted surgery, ventilation, and diet were part of the patient's health condition but the prolonged bradycardia was due to the use of the catheter.
 
Additional Manufacturer Narrative
Correction d4: 30147513.The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 17-aug-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
All information reasonably known as of 27-jul-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30147513, was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.The returned product was evaluated and no issues were observed on device.The root cause could not be determined since the reported failure was not reproduced in lab during sample evaluation.All information reasonably known as of 28-sep-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD CLOSED SUCTION SYSTEM FOR NEONATES / PEDIATRICS, 5 F, Y-ADAPTER
Common Device Name
VAP CLOSED SUCTION CATHETERS & ACCESSORIES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key14760384
Report Number8030647-2022-00160
Device Sequence Number1294371
Product Code BSY
UDI-Device Identifier00609038983837
UDI-Public00609038983837
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/21/2025
Device Model Number195-5
Device Catalogue NumberN/A
Device Lot Number30147513
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2022
Supplement Date Received by Manufacturer07/01/2022
07/20/2022
09/06/2022
Initial Report FDA Received Date06/21/2022
Supplement Report FDA Received Date07/27/2022
08/17/2022
09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age18 DA
Patient SexMale
-
-