• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SPYGLASS RETRIEVAL BASKET; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SPYGLASS RETRIEVAL BASKET; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Model Number M00546550
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter facility name: kyushu central hospital of the mutual aid association of public school teachers.Initial reporter city fukuoka-shi, fukuoka prefecture.(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a spybasket was used in the common bile duct during an cholangioscopic lithotripsy procedure performed on (b)(6) 2022.During the procedure, the basket failed to open when attempting to grasp a stone.The procedure took a longer time than expected; therefore, the physician aborted the procedure and decided to try again next time.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block e1: initial reporter facility name: (b)(6) hospital.E1: initial reporter city (b)(6).Block h6: impact code f1001 is being used to capture the reportable issue of aborted/cancelled procedure.Block h10: the returned spybasket was analyzed, and a visual inspection observed the working length presented several kinks.Functional inspection found the basket could be extended and retracted without any problems.The reported event of basket failure to open was not confirmed.Based on all available information, the working length kinked may be related to user manipulation and/or some technique applied during the procedure that may have affected the working length as observed.Therefore, the most probable root cause for the problem found is adverse event related to procedure.The cause code of the reported event of basket failure to open is no problem detected.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a spybasket was used in the common bile duct during an cholangioscopic lithotripsy procedure performed on (b)(6) 2022.During the procedure, the basket failed to open when attempting to grasp a stone.The procedure took a longer time than expected; therefore, the physician aborted the procedure and decided to try again next time.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYGLASS RETRIEVAL BASKET
Common Device Name
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14760496
Report Number3005099803-2022-03341
Device Sequence Number17668928
Product Code PTS
UDI-Device Identifier08714729965886
UDI-Public08714729965886
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/10/2023
Device Model NumberM00546550
Device Lot Number0026788051
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2022
Supplement Date Received by Manufacturer06/30/2022
Initial Report FDA Received Date06/21/2022
Supplement Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-