• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 100/189/070
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the hospital experimented on the product after cases of micro aspirations, and found that cuffs were leaking.The functionality tests were performed in the presence of a smiths rep and it was observed that inflated cuffs showed leaks and bubbles when immersed in transparent container filled with water.Additionally, it was alleged the spring valve of the pilot balloon wasn?t working right.The test was done in 189 and 199.They blew the cuff with air at 30 cm as recommended on the tube, then detached the manometer.They reconnected the manometer and after a second and observed that the reading dropped 26cm in size 7 and 8 et.The pressure was not getting stabilized and was continuously dropping.This is the same thing they are experiencing at patient bedside thought they were maintaining the cuff pressure for 3 times a day and still having vap case.Also tried to use 199 plain et and found the pressure was not getting stable, continuously dropping.They used 10 units to assess the functionality with the same outcome.No patient injury was reported.
 
Additional Manufacturer Narrative
Other, other text: device evaluation: no product was returned.We are unable to confirm the reported complaint.If the product is returned, smiths medical will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
thousand lights,
minneapolis, MN 55442
MDR Report Key14760516
Report Number3012307300-2022-12320
Device Sequence Number6106748
Product Code BTR
UDI-Device Identifier15019315059520
UDI-Public15019315059520
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K081086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/03/2026
Device Model Number100/189/070
Device Catalogue Number100/189/070
Device Lot Number4103881
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2022
Initial Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-