| Model Number |
DXC 700 AU-10DE, CHEMISTRY ANALYZER DXC 700 AU FOR DIRECT TR |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Beckman coulter field service engineer (fse) evaluated the dxc 700 au chemistry analyzer, and identified the failure mode as a defective sample syringe.The fse replaced the sample syringe to resolve the issue.Information not provided by customer.The beckman coulter internal identifier is (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported the generation of erroneous patient results for sodium (na), potassium (k), creatinine (cre) and cholesterol (chol) on their dxc 700 au chemistry analyzer.The results were not released outside of the laboratory.There was no report of change to treatment, injury, or death associated with this event.The customer did not provide patient demographics or data.
|
| |
|