The device was not returned to stryker sustainability solutions for evaluation.A review of the dhr could not be performed as the lot number and/or serial number was not reported.(b)(4) , fall alarm sell sheet, states that item number 8903el (chair sensor pad 30 days 8' long cord) and item number m10-400 (non-slip padding 12¿-144¿) are separate item numbers and are sold separately.The reported event could be attributed to the following: user expectation.Accessory not properly placed by user.The instructions for use (ifu) state: to reduce the risk of serious injury, always follow the directions for testing sensor pad (see below) after putting the sensor pad in place and before leaving patient unattended.Do not use any alarm or sensor pad that does not alarm each time it is tested.Use of restraints: the position of the sensor pad is vital when using a restraint.Make sure the restraint is applied correctly per instructions for that device.Straps must not cross over sensor pad.If straps cross over sensor pad and patient moves, pressure from straps may prevent alarm from activating.If patient falls out of chair and is suspended in the restraint, serious injury may occur from chest compression or suffocation.Use extreme caution with chair cushions.If a cushion dislodges, straps may loosen and allow patient to slide off seat and become suspended.Directions for use.Read all sensor pad and alarm instructions prior to use.Check that sensor pad, cord, and plug are clean and undamaged.If you have a damaged sensor pad and/or plug, the sensor pad unit should be discarded and replaced.When putting sensor pad into use, record the first use date and calculate expiration date from initial use.Place sensor pad flat across width of seat, labeled side up never roll, fold, or crease sensor pad.Make sure sensor pad cord is either to the back or side of seat and is not folded back under the sensor pad.If a seat cushion is used, place sensor pad on top of cushion.Adjust so sensor pad is directly under patient¿s buttocks when seated.Sensor pad should be towards front of chair seat if patient normally sits toward front.Sensor pad should be towards back of chair seat if using a posture support or if patient is at risk of forward sliding.If needed, use an incontinence pad to prevent damage from urine or other liquids.If liquid enters at ¿neck¿ of sensor pad, alarm may not function.Route the sensor pad cord to the posey® alarm.Check that the sensor pad cord has slack, is not stressed, is clear of moving parts of chair, and does not pose a tripping hazard.Insert plug into sensor input on the alarm.Check that the plug on the sensor pad cord is not damaged (plug broken, or wires disconnected) and is securely connected to the alarm.Test sensor pad and alarm (refer to directions for testing sensor pad).The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
|