STRYKER ENDOSCOPY-SAN JOSE PKG, STRYKERFLOW ELECTROCAUTERY PROBE, L TIP, 5MM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0250070460
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device arced.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
Alleged failure: can you enter a complaint, issue rma/pi# and send angie 2 x 250-070-460 as replacements? see notes below.There was no patient harm.The l tip was arcing and looks like sheath was compromised.Salesforce case number (b)(4) the failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be 1) conductive irrigating or aspirating fluid used was inside the distal end, and contacted the probe and the sheath causing the sheath to melt, 2) power set too high.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacturing date is unknown.
 
Event or Problem Description
It was reported that the device arced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, STRYKERFLOW ELECTROCAUTERY PROBE, L TIP, 5MM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key14762345
Report Number0002936485-2022-00378
Device Sequence Number6127726
Product Code GEI
UDI-Device Identifier07613327051872
UDI-Public07613327051872
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K963765
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0250070460
Device Catalogue Number0250070443
Device Lot Number050621-06
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/08/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date06/21/2022
Supplement Report FDA Received Date10/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-