Model Number M0035421030
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The device was returned and the reported lot number was confirmed from the packaging label.During visual/microscopic inspection, the coil delivery wire was found to be kinked/bent, the main coil was found to be broken/fractured, the main coil was found to be kinked/bent and the proximal contact wire was found to be broken/fractured approx.1 cm from the distal end.Coil in catheter friction functional test was unable to be performed due to breakage.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event of coil in catheter friction could not be confirmed; however, the analysis results are consistent with the reported event.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The device was analyzed and the main coil was found to be kinked/bent and broken/fractured.This complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.Therefore, an assignable cause of procedural factors will be assigned to the reported event of coil in catheter friction, and to the analyzed events of main coil kinked/bent, and main coil broken/fractured during use.An assignable cause of 'handling damage' will be assigned to analyzed event of coil delivery wire kinked/bent and coil proximal contact broken/fractured as these defects appear to be associated with handling of the product or portion of the product during the procedure.
 
Event or Problem Description
Analysis of the returned device found the main coil broke/fractured during use.There were no clinical consequences to the patient reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1MM X 3CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14764320
Report Number3008881809-2022-00306
Device Sequence Number17680456
Product Code HCG
UDI-Device Identifier04546540697974
UDI-Public04546540697974
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/24/2024
Device Model NumberM0035421030
Device Catalogue NumberM0035421030
Device Lot Number23062130
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/31/2022
Initial Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SL-10 MICROCATHETER (STRYKER)
Patient Age66 YR
Patient SexFemale
Patient Weight58 KG
Patient RaceAsian
-
-