Model Number MDCONS100
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was received and evaluated.Device inspection and evaluation, service repair noted the following : externally, the unit is in good condition.Internally, the unit is in good condition.When tested, the unit passed all tests in accordance with the oem (original equipment manufacturer) technical manual.The unit.Was soak tested for 3 days and no error occurred.Service repair noted the customer specified fault unconfirmed.When tested, the unit passed all tests in accordance with the oem technical manual.The unit was soak tested for 3 days and no error occurred.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event or Problem Description
It was reported that the device cuts out in mid use.The issue occurred during a therapeutic fess (functional endoscopic sinus surgery).According to the doctor (user), the operation was abandoned as the machine cuts out and could not be used and that the only problem is that the patient will need to return to have the procedure done again at another date.The reporter reported no patient harm or injury and no medical intervention reported as a result of the event.No user injury reported.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been more than 7 years since the device was manufactured.Based on the results of the investigation, it is likely that the reported phenomenon was caused by user error of the device.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIDEBRIDER POWER CONSOLE
Common Device Name
MULTIDEBRIDER POWER CONSOLE
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
Manufacturer Contact
brian motter
2925 appling rd
bartlett, TN 38133
4848965250
MDR Report Key14767136
Report Number1037007-2022-00011
Device Sequence Number14978947
Product Code GEI
UDI-Device Identifier00821925027770
UDI-Public00821925027770
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K123429
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMDCONS100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date06/22/2022
Supplement Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-