| Model Number |
HLO52535HK |
| Medical Device Problem Codes |
Device Contamination with Body Fluid (2317); Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiration date: 11/2023).
|
| |
|
Event or Problem Description
|
|
It was reported that prior to an angioplasty procedure, the device allegedly had plastic particles on the sheath.The procedure was completed using another device.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a complaint history review was performed.This is the first complaint reported for this product/lot number combination.However, the device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned for evaluation.The result of the investigation is unconfirmed for the reported plastic particle contamination issue but confirmed for biological contamination.The returned halo device did not exhibit the presence of any plastic particles.Solidified yellow residue (likely bodily substance via user transfer) was present on the sheath shaft and tip.The root cause for the reported plastic particle contamination issue could not be determined based upon the available information received from the field communication and sample evaluation.Labeling review: instruction for use for the halo one thin-walled guiding sheath was reviewed and the following sections are applicable: warnings 4.Visually inspect the packaging to verify that the sterile barrier is intact.Do not use if the sterile barrier is opened or damaged.Precautions: 5.The pouch should be inspected prior to opening to ensure the sterile barrier is not compromised.The device should be carefully removed and placed in the sterile field.The entire procedure from skin puncture or incision to sheath withdrawal must be carried out aseptically.6.Carefully inspect the sheath prior to use to verify that the sheath has not been damaged during shipment and that its size, shape and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.H10: d4 (expiration date: 11/2023), g3, h6(device).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to an angioplasty procedure, the device allegedly had plastic particles on the sheath.The procedure was completed using another device.There was no patient contact.
|
| |
|