U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 HEMOSPLIT CATHETER; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 HEMOSPLIT CATHETER; DIALYSIS CATHETER Back to Search Results
Model Number 5743730
Medical Device Problem Codes Collapse (1099); Material Discolored (1170); Material Opacification (1426); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiration date: 06/2022).
 
Event or Problem Description
It was reported that approximately two months post dialysis catheter placement procedure in the right internal jugular vein, the catheter allegedly got thin and collapsed in the left posterior organ cavity.It was further reported that the catheter was removed and replaced.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.One electronic photo was provided for review.The investigation is confirmed for reported deformation, and collapse and the identified discoloration issues, as the catheter was noted to be deformed longitudinally and discoloration were evident in the provided photo.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 06/2022), g3, h6 (method).H11: h6 (device, result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that approximately two months post dialysis catheter placement procedure in the right internal jugular vein, the catheter allegedly got thin and collapsed in the left posterior organ cavity.It was further reported that the catheter was removed and replaced.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPLIT CATHETER
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14768676
Report Number3006260740-2022-02360
Device Sequence Number6153316
Product Code MSD
UDI-Device Identifier00801741013287
UDI-Public(01)00801741013287
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K030020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5743730
Device Catalogue Number5743730
Device Lot NumberREEX2243
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/02/2022
Supplement Date Received by Manufacturer09/06/2022
Initial Report FDA Received Date06/22/2022
Supplement Report FDA Received Date09/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-