| Model Number |
ACU0T0 |
| Medical Device Problem Codes |
Defective Device (2588); Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported with a description of faulty intraocular lens (iol) was reported, additional information has been received and stated that, following an intraocular lens (iol) implant procedure, on evaluation of the eye post-operatively, it was noted that the lens has shifted from the bag due to faulty haptic and there was no longer in position.One month of initial implant the iol was repositioned successfully.The physician mentioned that, the issue is with the injector, not the lens itself.The lens come out twisted from the injector almost all the time.
|
| |
|
Additional Manufacturer Narrative
|
|
A non-qualified viscoelastic was indicated.The root cause may be related to a failure to follow the ifu (instructions for use).A non-qualified viscoelastic was used in the device.The ifu instructs: during device preparation and implantation of the company iq iol with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.The manufacturer internal reference number is: (b)(4).
|
| |
|