• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION DESTINATION RENAL GUIDING SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION DESTINATION RENAL GUIDING SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that during a peripheral arterial disease (pad) procedure via right leg access from the left common femoral artery (cfa), the destination device involved was advanced up and over a very calcified and steep bifurcation.The procedure was completed but upon removal of the device, the sheath skin tore exposing the coil inside.Hemostasis was achieved with an 8fr angio-seal.The procedure was completed.The event occurred intra-operative.The estimated blood loss was less than 250cc.The patient was in stable condition.The procedure outcome was completed.
 
Additional Manufacturer Narrative
Date of birth: requested, not provided.Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation: registered technician (rt).The actual device has been returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot number combination was conducted with no findings.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to update section h3, and to provide the completed investigation results.One 7fr destination sheath was returned for evaluation.The sheath was separated with the coil holding the two separated pieces together.The separation was measured to occur 30.5cm from the hub.The sheath had two pinch points, one measured to be 2.6cm from the tip, and the other 9.5 cm from the tip.The complaint can be confirmed for sheath separation based on the condition of the returned device.The exact root cause could not be determined.The likely root cause is that withdrawing the sheath over the very calcified and steep bifurcation caused the sheath to separate.Review of the device history record (dhr) showed there were no issues noted during manufacture of the product that could have contributed to this complaint condition.Currently no action is recommended since this risk evaluation is within the predetermined limits in the design failure mode and effects analysis (dfmea).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESTINATION RENAL GUIDING SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer (Section G)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer Contact
stephanie handy
950 elkton blvd.
elkton, MD 21921
9499890491
MDR Report Key14770828
Report Number1118880-2022-00034
Device Sequence Number1334833
Product Code DYB
UDI-Device Identifier00389701006192
UDI-Public00389701006192
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
K172995
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2024
Device Model NumberN/A
Device Catalogue NumberRSR04
Device Lot Number0000157837
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date06/22/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8FR ANGIO-SEAL
Patient SexFemale
Patient Weight54 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-